Shenzhen Yehai Sci. & Tech. Deve. Co., Ltd.
Products
Contact Us
  • Contact Person : Ms. Chow Tracy
  • Company Name : Shenzhen Yehai Sci. & Tech. Deve. Co., Ltd.
  • Tel : 86-755-26509606-8803
  • Fax : 86-755-26502620
  • Address : Guangdong,Shenzhen,Room No.216,Third Floor,Jin Huiqiu Building,Langshan No.2Road,NorthArea,Science&TechnologyPark,Nanshan District,Shenzhen,China
  • Country/Region : China
  • Zip : 518057

Loop CalibratorYH7007 with Rapid induction rotating

Loop CalibratorYH7007 with Rapid induction rotating
Product Detailed
Loop Calibrator YH7007 OEM Servicera Max voltage 30V Large display,simple, quick-click interface Fluke 707 Loop Calibrator

Loop Calibrator-YH7007

We can provide OEM Service for customers completely

Features

The fast, one-handed tool for loop checks. Just click to calibrate.

This feature-rich solution combines the best features push-button and knob-base loop calibrator and offers:

Large display and simple, quick-click interface for easy one-handed operation.Simultaneous mA and % readout for quick, easy interpretation of readings.mA accuracy of 0.025%, superior to other loop calibrators in this price range.1 µA resolution for mA source, simulate and measure.Push button 25% steps for fast, easy linearity checks.0-100% “span check” for fast confirmation of zero and span.Selectable slow ramp, fast ramp, and step ramp to provide smooth outputs for valve slewing and loop functional tests. 24 volt internal loop supply, so you can power and read a transmitter at the same time without carrying a DMM.Measures V dc to 28 V.Single 9V battery that is easily changed without braking calibration seal0 - 20 mA or 4 - 20 mA default start up modes.Innovative output adjustment dial with 1 µA and 100 resolution.HART ™ mode connects 250 Ω resistor in series with 24 V loop for compatibility with HART communicators.Specifications

Current Measurement

Range:

0 to 24 mA

Resolution:

0.001 mA

Accuracy:

0.025% Rdg + 2 LSD

Current Sourcing

Range:

0 to 20 mA or 4 to 20 mA

Accuracy:

0.025% Rdg + 2 LSD

Drive Capability:

1200 Ω; 950 Ω at 20 mA in HART mode

Current Simulation

Range:

0 to 20 mA or 4 to 20 mA

Accuracy:

0.025% Rdg + 2 LSD

Ext. loop voltage requirement:

12-30 V

Loop Power

While measuring mA, 24 V

Voltage measurement

Range:

0 to 28 V DC

Resolution:

1 mV

Accuracy:

0.025% Rdg + 2 LSD

Environmental Specifications

Operating Temperature -10ºC to 55ºC

Storage Temperature -30ºC to 60ºC

Humidity (Without Condensation) 0 to 90% (0ºC to 35ºC); 0 to 70% (35ºC to 55ºC)

Temperature Coefficient -10ºC to 18ºC, 28ºC to 55ºC, ±.005% of range per ºC

Safety Specifications

Maximum voltage 30V

Mechanical & General Specifications

Size 152 x 70 x 41 mm

Weight 224 g (340 g including holster)

Batteries One 9V battery Alkaline(YHS-707) , One 9V battery

Warranty Two years

Shock & Vibration per MIL-T-28800 for a Class 2 Instrument

Display current and % of span Yes

Auto step, auto ramp Yes

Span Check Yes

Loop CalibratorYH7007 with Rapid induction rotating



Copyright Notice @ 2008-2022 ECBAY Limited and/or its subsidiaries and licensors. All rights reserved.