- PH Meters[3]
- Flow Meters[1]
- Temperature Instruments[10]
- Resistance Meters[10]
- Clamp Meters[10]
- Multimeters[10]
- Used Measuring & Analysing Instruments[2]
- Testing Equipment[10]
- Voltage Meters[5]
- Other Electrical Instruments[10]
- Laser Rangefinders[2]
- Other Physical Measuring Instruments[1]
- Other Electrical Equipment[3]
- Signal Generators[3]
- Measuring & Analysing Instrument Processing Services[2]
- Pressure Transmitters[2]
- Industrial Metal Detectors[1]
- Other Measuring & Analysing Instruments[10]
- Measuring & Analysing Instrument Design Services[1]
- Speed Measuring Instruments[1]
- Other Analysis Instruments[1]
- Level Measuring Instruments[1]
- Current Meters[1]
- Power Meters[5]
- Contact Person : Ms. Chow Tracy
- Company Name : Shenzhen Yehai Sci. & Tech. Deve. Co., Ltd.
- Tel : 86-755-26509606-8803
- Fax : 86-755-26502620
- Address : Guangdong,Shenzhen,Room No.216,Third Floor,Jin Huiqiu Building,Langshan No.2Road,NorthArea,Science&TechnologyPark,Nanshan District,Shenzhen,China
- Country/Region : China
- Zip : 518057
5Tohm; 5KV Digital Insulation Tester YH510
Features
. Checking the insulation of high-voltage electrical equipment
(electric motor, transformer, and cable, etc)
.Test Voltage from 250V~5000V
.Insulation Resistance up to 5Tohm
. Short Current up to 3mA
. Auto Calculate PI(Polarization Index) and DAR(Dielectric absorption ratio)
. Step-Voltage Measurement
. Temperature measurement
. Temperature compensation for insulation Test
. Leakage Current measurement
. Data Storage function and uploading to PC by USB2.0 interface
. Power Supply:LR6 Battery×6/ rechargeable battery Ni~MH 7.2V
(Charger;Input Voltage 100V~240V ; Frequency 50Hz~60Hz, Output Voltage 12V DC 3A)
. Timer Function
.Weight 2.5kg (not including Battery)
.Size 284mm(L)×230mm(w)×125mm(H)
Test Voltage
. Test Voltage Range; DC250V~ DC 5.00kv
. Voltage Set:
. Wide adjust;(250V; 500V; 1000V; 2.5kv; 5kv)
. Fine adjust (250V~1KV; Step 25V; 1KV~5KV, Step 100V)
. Accuracy; 5%
. Step mode:500V, 1.0KV, 1.5KV, 2.0KV, 2.5KV
. Test Current 1nA~3mA
Temperature compensation
For obtaining the insulation resistance under a temperature other than current ambient temperature.
Converts the measured value of resistance to the insulation resistance under a reference temperature, and displays the result.
Depending on the different properties of test objects, there are 10 modes of compensation.
Insulation Resistance
Set test voltage | Test range | Threshold value of basic error |
250 V DC | 0.01 Mohm ~ 2.5 Gohm | ±(5% rdg + 5 dgt) |
2.51 Gohm ~ 250 Gohm | ±(20% rdg +10 dgt) | |
500 V DC | 0.01 Mohm ~ 5.0 Gohm | ±(5% rdg +5 dgt) |
5.01 Gohm ~ 500 Gohm | ±(20% rdg +10 dgt) | |
1000 V DC | 0.01 Mohm ~ 10 Gohm | ±(5% rdg +5 dgt) |
10.1 Gohm ~ 500 Gohm | ±(20% rdg +10 dgt) | |
501 Gohm ~ 999 Gohm | ±(30% rdg +20 dgt) | |
2.5 kV DC | 0.01 Mohm ~ 25 Gohm | ±(5% rdg +5 dgt) |
25.1 Gohm ~ 500 Gohm | ±(20% rdg +10 dgt) | |
501 Gohm ~ 999 Gohm | ±(30% rdg +20 dgt) | |
1 Tohm ~ 2.5 Tohm | ±(30% rdg + 40 dgt) | |
5 kV DC | 0.01 Mohm ~ 50 Gohm | ± (5% rdg + 5 dgt) |
50.1 Gohm ~ 500 Gohm | ±(20% rdg + 10 dgt) | |
501 Gohm ~ 999 Gohm | ±(30% rdg + 20 dgt) | |
1 Tohm ~ 5 Tohm | ±(30% rdg + 40 dgt) |
5Tohm; 5KV Digital Insulation Tester YH510